Accelerating voltage scan: Difference between revisions
(→Orange Book Entry: change case in heading, Replaced: Book Entry == ‚Äö√ú√≠ Book entry ==, using AWB) |
(Orange Book cat case, Replaced: == Orange Book entry == ‚Üí <!-- Orange Book -->, == Gold Book entry == ‚Üí <!-- Gold Book -->, using AWB) |
||
(3 intermediate revisions by 2 users not shown) | |||
Line 2: | Line 2: | ||
<!-- Orange Book --> | |||
{{orange| | {{orange| | ||
A method of producing a momentum (mass) spectrum in a magnetic deflection instrument by varying the accelerating voltage. This scan can also be used in conjunction with a fixed radial electric field to produce an ion kinetic energy spectrum. | A method of producing a momentum (mass) spectrum in a magnetic deflection instrument by varying the accelerating voltage. This scan can also be used in conjunction with a fixed radial electric field to produce an ion kinetic energy spectrum. | ||
}} | }} | ||
<!-- Gold Book --> | |||
{{gold| | |||
http://goldbook.iupac.org/A00050.html | |||
This is an alternative method of producing a momentum (mass) spectrum in magnetic deflection instruments by varying the accelerating voltage. This scan can also be used, in conjunction with a fixed radial electric field, to produce an ion kinetic energy spectrum. | |||
Source: PAC, 1991, 63, 1541 ([[Recommendations for nomenclature and symbolism for mass spectroscopy]] (including an appendix of terms used in vacuum technology). (Recommendations 1991)) on page 1550 | |||
}} | |||
[[Category:M/z Separation]] | [[Category:M/z Separation]] | ||
{{DEFAULTSORT:Accelerating Voltage Scan}} | {{DEFAULTSORT:Accelerating Voltage Scan}} | ||
[[Category:Orange Book term]] |
Latest revision as of 11:50, 18 July 2009
The Accelerating voltage scan page currently does not have any content, please see |
All terms | IUPAC 2013 | Gold Book terms | Orange Book terms | Acronyms | Deprecated terms | Obsolete terms | Draft definitions | Stub pages |
Orange Book
ORANGE BOOK DEFINITION
IUPAC. Analytical Division. Compendium of Analytical Nomenclature (the Orange Book). Definitive Rules, 1979 (see also Orange Book 2023) |
Accelerating voltage scan |
---|
A method of producing a momentum (mass) spectrum in a magnetic deflection instrument by varying the accelerating voltage. This scan can also be used in conjunction with a fixed radial electric field to produce an ion kinetic energy spectrum. |
IUPAC 1997 Orange Book Chapter 12 |
Index of Orange Book Terms |
Gold Book
GOLD BOOK DEFINITION
IUPAC. Compendium of Chemical Terminology, 2nd ed. (the Gold Book). Compiled by A. D. McNaught and A.Wilkinson. Blackwell Scientific Publications, Oxford (1997). |
Accelerating voltage scan |
---|
http://goldbook.iupac.org/A00050.html This is an alternative method of producing a momentum (mass) spectrum in magnetic deflection instruments by varying the accelerating voltage. This scan can also be used, in conjunction with a fixed radial electric field, to produce an ion kinetic energy spectrum. Source: PAC, 1991, 63, 1541 (Recommendations for nomenclature and symbolism for mass spectroscopy (including an appendix of terms used in vacuum technology). (Recommendations 1991)) on page 1550 |
IUPAC Gold Book |
Index of Gold Book Terms |